Abstract
Objective: Today, different trademarks of Hematology Analyzers have been used in blood analysis and but the compatibility of these devices with each other is unknown. In this study, the aim was to investigate the agreement between the results of complete blood countparameters values with BC 6000 hematology analyzer (Mindray, China) and BT PRO 2401 hematology analyzer (Izmir, Turkey). Materials and Methods: This prospective study was included in 51 EDTA-anticoagulated samples submitted to our biochemistry laboratory for routine testing of CBC(complete blood cell). Statistical correlations CBC parameters values, analyzed in each analyzer as a single analysis from each sample were evaluated using two-way random intraclass correlation coefficient (ICC) and regression analysis. We compared the accuracy and precision of CBC count methods; i.e. BC-6000 Plus (optical (O) and impedance (I))and BT PRO 2401 (optical (O) and impedance (I)). Results: There was agreement between the two devices in Wbc, Neu, Mon, Neuperc, Lymperc, Monperc, Hct, Mcv, Mch, RdwSD, Plt, Mpv and pct parameters as intraclass correlation coefficient (ICC) <90. However we found disagreement in Bas, Eos, Eosper, Basper, Rbc, Hgb,Mchc hemoglobin, RdwCV and Pdw measurements between Mindray and Bt-pro 2401 (ICC >90). Conclusion:In terms of white blood cells and neutrophil analysis which suggests that they can be used interchangeably, perfect agreement between them. However, instruments did not ensure satisfactory interchangeability and did not facilitate